The Intempco MIST PKIT-1 Programming Module and Software is used to
re-calibrate and re-scale MIST Series sensors. Two methods of calibration
are available. The first, One Point Calibration, is performed by setting
the MIST to a known temperature and adjusting the 4-20mA output to the
corresponding value.
The second method, Two Point Calibration, is performed by setting the
MIST to a known point in the operating range and then recording this point
by simply clicking the corresponding button. The next step is to set the
MIST to a second known point in the operating range and record it in the
same manner. Finally, by clicking on the “Process” button, the calibration
parameters are automatically calculated and saved in the MIST memory.
The MIST software can also be used to re-scale the MIST sensor output
within its operational limits. A choice of many pre-scaled values is
offered, as well as a custom range. Once the range is selected, simply
clicking on the “Set” button will automatically re-scale the output of the
MIST and save the new values in the MIST memory.
The Intempco MIST PKIT-1 Programming Module and Software kit includes
calibration software on a CD-ROM, a communication module with RS-232, and
a 110 VAC wall adapter.
Minimum System Requirements:
- Windows™ 95, 98, NT 4.0, Millenium, 2000 or XP
- Pentium 1 - 300 MHZ or better
- 5 MB of free memory
- 16 MB of RAM
- 1 serial port
This product is an accessory for the following products: |
Quick Specs
-
Part number MIST PKI-1
-